PCTM 4B
ISSUED-
1996
The information and data contained
in this document were prepared by a technical committee of the Association. The
committee and the Association assume no liability or responsibility in connection
with the use of such information or data, including but not limited to any
liability responsibility under patent, copyright, cc trade secret laws. The
user is responsible for determining that this document is the most recent
edition published.
Moisture
content of tall oil by coulometry
Scope
This test method covers the
determination of water in liquid organic materials containing <2% water. The
moisture content of tall oil is determined by automatic titration with coulometrically
generated Karl Fisher reagent In the Karl Fisher reaction, iodine reacts with
water in the presence of sulfur dioxide, an organic nitrogen base like pyridine
and an alcohol:
I2+ H2O
+ SO2+
ROH + 3RN→2RN•H1+
RN•HSO4R
With coulomettic titrators the
required iodine is produced by anodic oxidation of an iodide contained in the
cell reagents. The sample water content is then determined directly by
electronically integrating the current required to generate sufficient iodine
to precisely react with the water in the sample.
Safety precautions
Always operate the titrator in a
fume hood. Most of the reagents, including some of the pyridine-free systems
are noxious or toxic and inhalation or direct skin contact with them should be
avoided.
Apparatus
1. Coulometric Karl Fisher titrator.
2. Sample vials, 8-nt or 4-dram.
3. Medicine droppers.
4. Disposable plastic syringes, 1-cc
with 16, 18 or 20 gauge needles.
5. Syringe, 10-uL.
Reagents
Karl Fischer pyridine-free reagent.
Reagents are available from the manufacturer of the apparatus.
Instrument preparation
Prior to sample analysis the analyst
must become familiar with the operation of the titrator which differs slightly depending
on the manufacturer of the apparatus.
1.Prepare and calibrate the
apparatus as specified in the operating manual supplied by the manufacturer.
This generally includes the following step
a.
Clean and assemble the titration chamber.
b.Pour
the titration solution into the chamber.
c.Add
the iodine generator solution to the generator assembly.
d.Put
the instrument in STANDBY position.
e.Slowly
add neutralizing solution normally by injection with a syringe)
f.Continue the addition until the instrument indicates by a
light or a message on an LCD that excess water is present.
g.Stop
the addition of neutralizing solution.
h.Let
the instrument stabilize for 1 hour in the STANDBY position.
2. Verify the instrument by
injecting a known amount of water, usually 1 uL to 5 uL below the surface of
the titration solution.
3. Put the instrument in the TITRATE
or RUN position and record the reading when the END light comes on or a message
appears that the titration is complete.
Property of the Pine Chemicals
Association. All Rights Reserved.
Reproduction Restricted to Personal
Use.
Tidak ada komentar:
Posting Komentar